Remote Control Commands
R&S
®
SMW200A
792User Manual 1175.6632.02 ─ 16
Example:
// Start adjustments for maximum frequency accuracy
CALibration:FREQuency:MEASure?
// 0
// Adjustments are performed successfully
Usage: Query only
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency and baseband
gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q imbalance and
quadrature.
Return values:
<Local> 0 | 1 | OFF | ON
*RST: 0
Example:
// Start adjustment for the I/Q modulator for the currently set frequency
CALibration:IQModulator:LOCal?
// 0
// Adjustments are performed successfully
Usage: Query only
Manual operation: See "Adjust I/Q Modulator Current Frequency" on page 419
:CALibration<hw>:LEVel:HACCuracy[:STATe] <State>
Activates level adjustment with high accuracy.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
CALibration:LEVel:HACCuracy:STATe 1
:CALibration<hw>:FREQuency:CONVerter:EXTernal?
Queries the calibration state of the external I/Q upconverter R&S SZU.
Return values:
<Success> 0 | 1 | OFF | ON
*RST: 0
Example:
CAL1:FREQ:CONV:EXT?
// 1
// The R&S SZU connected in path A is calibrated.
Usage: Query only
Manual operation: See "Adjust External Frequency Converter" on page 1162