Remote Control Commands
R&S
®
SMW200A
810User Manual 1175.6632.02 ─ 16
Parameters:
<AMode> FIXed | MANual | AUTO
AUTO
The attenuator adjusts the level settings automatically, within the
full variation range.
FIXed
The attenuator and amplifier stages are fixed at the current posi-
tion, providing level settings settings with constant output
VSWR. The resulting variation range is calculated according to
the position.
To use this mode, activate the ALC (see [:SOURce<hw>]:
POWer:ALC[:STATe]).
MANual
You can set the level manually, in 10 dB steps.
*RST:
AUTO
Example:
SOURce:POWer:ALC:STATe 1
OUTPut:AMODe FIXed
Manual operation: See " Mode " on page 523
:OUTPut<hw>:BLANk:LIST:STATe <State>
Activates RF output blanking.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
OUTP:BLAN:LIST:STAT ON
Activates the RF output blanking.
:OUTPut<hw>:IMPedance?
Queries the impedance of the RF outputs.
Return values:
<Impedance> G1K | G50 | G10K
*RST: G50
Example:
OUTP:IMP?
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage: Query only