Remote Control Commands
R&S
®
SMW200A
811User Manual 1175.6632.02 ─ 16
:OUTPut<hw>:AFIXed:RANGe:LOWer?
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the settable minimum/maximum value in mode :OUTPut:AMODe FIXed, i.e.
when the attenuator is not being adjusted.
See :OUTPut<hw>:AMODe on page 809
Return values:
<Upper> float
Increment: 0.01
Default unit: dBm
Example:
OUTPut1:AMODe FIXed
OUTPut1:AFIXed:RANGe:UPPer?
// -27
OUTPut1:AFIXed:RANGe:LOW?
// -50
Usage: Query only
Manual operation: See " Attenuator Level Range " on page 437
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped.
To define the output state, use the command :OUTPut<hw>[:STATe].
Example:
OUTP:PROT:CLE
Resets the protective circuit of the RF output.
Usage: Event
Manual operation: See "Overload" on page 524
:OUTPut<hw>:PROTection:STATe <State>
Attenuates the RF output signal for about 40 dB to protect external devices against
internal signals.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP:PROT:STAT ON
Attenuates the RF Output signal for about 40 dB.
Manual operation: See "Overload" on page 524
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
OUTPut Subsystem