Remote Control Commands
R&S
®
SMW200A
827User Manual 1175.6632.02 ─ 16
// Response: OUT
SCONfiguration:EXTernal:IQOutput1:DIRection?
SCONfiguration:EXTernal:CODer1:IQConnection:STATe?
// Response: 1
SCONfiguration:EXTernal:BBMM1:IQConnection:STATe?
// Response: 1
SCONfiguration:EXTernal:REMote:LIST?
// Response: SMBV100A,SGS_2,SGT100A,SMBV_1,SMBV_2,SMU200A (100010)
SCONfiguration:EXTernal:REMote:ADD "SGS_1","USB","100007"
SCONfiguration:EXTernal:REMote:ADD "SMU200A (100001)","LAN","10.112.11.125","2"
SCONfiguration:EXTernal:REMote:LIST?
// Response: SMBV100A,SGS_2,SGT100A,SMBV_1,SMBV_2,SMU200A (100010),SGS_1,SMU200A (100001)
SCONfiguration:EXTernal:REMote:REName "SGS_2","SGS (102030)"
SCONfiguration:EXTernal:REMote:LIST?
// Response:
// SMBV100A,SGS (102030),SGT100A,SMBV_1,SMBV_2,SMU200A (100010),SGS_1,SMU200A (100001)
// Detecting a connected external instrument and connecting it
SCONfiguration:EXTernal:BBMM1:REMote:DETect?
// Response: "SGT100A"
SCONfiguration:EXTernal:BBMM1:REMote:ISELect?
// Response: "SGT100A","A"
SCONfiguration:EXTernal:BBMM1:REMote:CONNect
// Manually defining the external instrument
SCONfiguration:EXTernal:IQOutput1:REMote:ISELect "SGS_1","A"
SCONfiguration:EXTernal:IQOutput1:REMote:CONNect
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
// Response: 1
SCONfiguration:EXTernal:BBMM1:RCONnection:STATe?
// Response: 1
SCONfiguration:EXTernal:CODer1:INAMe?
SCONfiguration:EXTernal:BBMM1:INAMe?
// Response: "SGT100A [A]"
SCONfiguration:EXTernal:IQOutput1:INAMe?
// Response: "SGS_1 [A]"
SCONfiguration:EXTernal:BBMM1:RF:COUPling ON
SCONfiguration:EXTernal:BBMM1:RF:FREQuency:OFFSet 0
SCONfiguration:EXTernal:BBMM1:RF:POWer:OFFSet -10
:OUTput1:STATe ON
SCONfiguration:EXTernal:BBMM1:RF:STATe?
// Response: 1
SCONfiguration:EXTernal:IQOutput1:RF:COUPling OFF
SCONfiguration:EXTernal:IQOutput1:RF:FREQuency 2143000000