828User Manual 1175.6632.02 ─ 16
SCONfiguration:EXTernal:IQOutput1:RF:POWer -20
SCONfiguration:EXTernal:IQOutput1:RF:STATe ON
SCONfiguration:EXTernal:REMote:INITialization:PREDefined:CATalog?
// Response: "external_reference_10MHz.iec"
SCONfiguration:EXTernal:IQOutput1:REMote:INITialization:PREDefined:FILE
"external_reference_10MHz"
SCONfiguration:EXTernal:IQOutput1:REMote:INITialization:FILE?
// Response: "external_reference_10MHz"
// Alternatively use an user defined initialization file
// MMEMory:CDIRectory "/var/user/RcExtInstr"
// SCONfiguration:EXTernal:REMote:INITialization:USER:CATalog?
// Response: init_seq_sgs
// SCONfiguration:EXTernal:IQOutput1:REMote:INITialization:USER:FILE "init_seq_sgs.iec"
SCONfiguration:EXTernal:ACONnect 1
*RST
SCONfiguration:EXTernal:BBMM1:INAMe?
SCONfiguration:EXTernal:REMote:PURGe
// Disabling remote control of the external instruments
SCONfiguration:EXTernal:REMote:DISConnect[:ALL]
Example: Configuring master-slave mode
// in the first (master) instrument
SOURce1:ROSCillator:SOURce INT
SCONfiguration:MULTiinstrument:MODe MAST
SCONfiguration:MULTiinstrument:STATe 1
SCONfiguration:APPLy
SOURce1:INPut:USER6:DIRection OUTP
OUTPut1:USER6:SIGNal INSinc
// in the second (slave) instrument
SOURce1:ROSCillator:SOURce EXT
SOURce1:ROSCillator:EXTernal:FREQuency 10MHZ
SCONfiguration:MULTiinstrument:MODe SLAV
SCONfiguration:MULTiinstrument:STATe 1
SOURce1:INPut:USER5:DIRection INP
SOURce1:INPut:USER5:SIGNal OUTSinc
SOURce1:INPut:USER6:DIRection OUTP
OUTPut1:USER6:SIGNal INSinc
SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?
// SYNC