Remote Control Commands
R&S
®
SMW200A
847User Manual 1175.6632.02 ─ 16
Usage: Query only
Manual operation: See "Instrument Name" on page 153
:SCONfiguration:EXTernal:FADer<ch>:RF:COUPling <RfCouplingState>
:SCONfiguration:EXTernal:BBMM<ch>:RF:COUPling <RfCouplingState>
:SCONfiguration:EXTernal:IQOutput<ch>:RF:COUPling <RfCouplingState>
Enables/disables coupling all major RF setting (like the frequency, level and RF state)
of the external instrument to the R&S SMW.
Parameters:
<RfCouplingState> 0 | 1 | OFF | ON
*RST: 1
Example: see Example "Connecting and configuring external instrument
(advanced configuration)" on page 826
Manual operation: See "RF Couple" on page 153
:SCONfiguration:EXTernal:FADer<ch>:RF:FREQuency <Frequency>
:SCONfiguration:EXTernal:BBMM<ch>:RF:FREQuency <Frequency>
:SCONfiguration:EXTernal:IQOutput<ch>:RF:FREQuency <Frequency>
In uncoupled mode, sets the RF frequency of the external instrument.
Parameters:
<Frequency> float
Range: 100E3 to 3E9
Increment: 0.01
*RST: 1E9
Example:
SCONfiguration:EXTernal:IQOutput1:RF:COUPling OFF
SCONfiguration:EXTernal:IQOutput1:RF:FREQuency 2143E6
// RF frequency of the external instrument is RF = 2.143 GHz
Manual operation: See "(Delta) RF Frequency/ RF Level" on page 154
:SCONfiguration:EXTernal:FADer<ch>:RF:FREQuency:OFFSet <FreqOffset>
:SCONfiguration:EXTernal:BBMM<ch>:RF:FREQuency:OFFSet <FreqOffset>
:SCONfiguration:EXTernal:IQOutput<ch>:RF:FREQuency:OFFSet <FreqOffset>
In coupled mode, offsets the RF frequency of the external instrument with the selected
delta value.
Parameters:
<FreqOffset> float
Range: -3E9 to 3E9
Increment: 0.01
*RST: 0
SCONfiguration Subsystem