Remote Control Commands
R&S
®
SMW200A
849User Manual 1175.6632.02 ─ 16
:SCONfiguration:EXTernal:FADer<ch>:RF:STATe <RemConnState>
:SCONfiguration:EXTernal:BBMM<ch>:RF:STATe <RemConnState>
:SCONfiguration:EXTernal:IQOutput<ch>:RF:STATe <RemConnState>
Queries/sets the RF output state of the connected external instrument.
Parameters:
<RemConnState> 0 | 1 | OFF | ON
Example:
:SCONfiguration:EXTernal:IQOutput1:RF:COUPling
ON
:OUTput1:STATe ON
:SCONfiguration:EXTernal:IQOutput1:RF:STATe?
Response: ON
Example:
:SCONfiguration:EXTernal:BBMM1:RF:COUPling OFF
:SCONfiguration:EXTernal:BBMM1:RF:STATe ON
Manual operation: See "RF State" on page 154
:SCONfiguration:EXTernal:REMote:SCAN
Scans the network for connected instruments, like R&S SMW, R&S SGS, R&S SMBV,
R&S SMU, R&S SMATE and R&S SMJ.
Example:
see Example "Connecting and configuring external instrument
(advanced configuration)" on page 826
Usage: Event
Manual operation: See "Detect/Scan" on page 156
:SCONfiguration:EXTernal:REMote:SCAN:STATe?
Queries if scanning is performed or not.
To start the scanning process, use the command :SCONfiguration:EXTernal:
REMote:SCAN on page 849.
Return values:
<ScanState> 0 | 1 | OFF | ON
1
Scanning process running
0
Not scanning
Example: See Example "Connecting and configuring external instrument
(advanced configuration)" on page 826
Usage: Query only
Manual operation: See "Detect/Scan" on page 156
SCONfiguration Subsystem