EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 888

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
873User Manual 1175.6632.02 ─ 16
Parameters:
<Source> LF1 | LF2 | NOISe | EXT1 | EXT2 | EXTernal | INTernal
LF1|LF2
Uses an internally generated LF signal.
EXT1|EXT2
Uses an externally supplied LF signal.
NOISe
Uses the internally generated noise signal.
INTernal
Uses the internally generated signal of LF1.
EXTernal
Uses an external LF signal (EXT1).
*RST:
LF1 <AM1>; LF2 <AM2>
Example: See Example "Creating an amplitude modulated RF signal"
on page 870.
Manual operation: See "Source" on page 507
[:SOURce<hw>]:AM<ch>:STATe <State>
Activates amplitude modulation.
Suffix:
<ch>
.
1 to 2
Modulation signal channel
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Creating an amplitude modulated RF signal"
on page 870.
Manual operation: See "State" on page 506
[:SOURce<hw>]:AM<ch>[:DEPTh] <Depth>
Sets the depth of the amplitude modulation in percent.
Suffix:
<ch>
.
1 to 2
Modulation signal channel
Parameters:
<Depth> float
Range: 0 to 100
Increment: 0.1
Example: See Example "Creating an amplitude modulated RF signal"
on page 870.
SOURce Subsystem