Remote Control Commands
R&S
®
SMW200A
873User Manual 1175.6632.02 ─ 16
Parameters:
<Source> LF1 | LF2 | NOISe | EXT1 | EXT2 | EXTernal | INTernal
LF1|LF2
Uses an internally generated LF signal.
EXT1|EXT2
Uses an externally supplied LF signal.
NOISe
Uses the internally generated noise signal.
INTernal
Uses the internally generated signal of LF1.
EXTernal
Uses an external LF signal (EXT1).
*RST:
LF1 <AM1>; LF2 <AM2>
Example: See Example "Creating an amplitude modulated RF signal"
on page 870.
Manual operation: See "Source" on page 507
[:SOURce<hw>]:AM<ch>:STATe <State>
Activates amplitude modulation.
Suffix:
<ch>
.
1 to 2
Modulation signal channel
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Creating an amplitude modulated RF signal"
on page 870.
Manual operation: See "State" on page 506
[:SOURce<hw>]:AM<ch>[:DEPTh] <Depth>
Sets the depth of the amplitude modulation in percent.
Suffix:
<ch>
.
1 to 2
Modulation signal channel
Parameters:
<Depth> float
Range: 0 to 100
Increment: 0.1
Example: See Example "Creating an amplitude modulated RF signal"
on page 870.
SOURce Subsystem