Remote Control Commands
R&S
®
SMW200A
894User Manual 1175.6632.02 ─ 16
Parameters:
<SRate> float
Range: 50 to depends on the installed options, e.g. 100E6
(R&S SMW-B10)
Increment: 1E-3
*RST: 270833.333
Example: See Example "Performing general tasks" on page 891
Manual operation: See "Symbol Rate" on page 267
[:SOURce<hw>]:BB:DM:STANdard
<Standard>
Selects predefined set of settings according to the selected standard, see Table 4-11.
Parameters:
<Standard> USER | BLUetooth | DECT | ETC | GSM | GSMEdge | NADC |
PDC | PHS | TETRa | W3GPp | TDSCdma | CFORward |
CREVerse | WORLdspace | TFTS | APCOPH1C4fm |
APCOPH1CQpsk | APCOPH2HCpm | APCOPH2HDQpsk |
APCOPH2HD8PSKW | APCOPH2HD8PSKN | APCOPH1Lsm |
APCOPH1Wcqpsk | CWBPsk
A query returns the value USER if one the following is true:
• A user-defined Custom Dig Mod setting was loaded
• One of the associated settings was changed subsequent to the
selection of a standard.
*RST: GSM
Example: See Example "Performing general tasks" on page 891
Options: GSM requires R&S SMW-B10
Manual operation: See "Set according to Standard" on page 267
[:SOURce<hw>]:BB:DM:SOURce <Source>
Selects the data source, e.g. a sequence of 0 or 1, a pseudo-random sequence with
different length, a pattern, a data list (DLISt), or external serial data (SERial).
Parameters:
<Source> ZERO | ONE | PRBS | PATTern | DLISt | SERial
*RST: PRBS
Example: See Example "Performing general tasks" on page 891
Options: SERial requires R&S SMW-B10
Manual operation: See "Data Source" on page 269
SOURce Subsystem