Getting Started
R&S
®
SMW200A
79User Manual 1175.6632.02 ─ 16
In the provided example, you use the R&S SMW to generate two baseband signals,
apply a baseband frequency shift, weight them and add them. You then route the gen-
erated stream and define the output connector. The initial situation is the configuration
described in Chapter 2.3.3, "Triggering the Instrument with an External Signal",
on page 71.
The minimum requirement for the instrument in this example is an R&S SMW equipped
with the options 2xR&S SMW-B10, R&S SMW-B13T, and 2xR&S SMW-B103.
To configure a composed baseband signal
1. In the first baseband, generate a WCDMA-3GPP signal as described in Chap-
ter 2.3.2, "Generating a Digitally Modulated Signal", on page 69.
2. To route the signal from "Baseband B" block to the first signal path, use one of the
alternatives:
In the block diagram, select "Baseband B" block and drag&drop it to the first
path ("Fading A" block or "AWGN A" block depending on the options installed
on your instrument).
Select the "Fading B" block and drag&drop it to the "Baseband B" block.
In the block diagram, select "Baseband B" block, navigate in the context menu,
and select "Signal Routing > route to path A".