Remote Control Commands
R&S
®
SMW200A
933User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:BB:ARBitrary:WAVeform:POINts?
Queries the number of samples (the number of I/Q values pairs) in the selected wave-
form file.
Return values:
<Points> <waveform filename>
Range: 0 to 1000
*RST: 1
Example: see Example "Managing waveform files" on page 920
Usage: Query only
[:SOURce<hw>]:BB:ARBitrary:WAVeform:DATA <Filename>, <Data>
[:SOURce<hw>]:BB:ARBitrary:WAVeform:DATA? [<Filename>, ]<Tag>
The setting command writes the binary block data <data> to the file identified by
<filename>. The complete content of the waveform file (i.e. including all tags) must
be specified; the complete content is transmitted as binary data block.
I/Q data and the marker data can also be written to a file with the command :
MMEMory:DATA:UNPRotected.
Tip: To ensure trouble-free data transmission, set the GPIB delimiter to EOI.
The query command retrieves the content of the specified tag of the currently selected
waveform file or the waveform file specified with the <filename>.
See also "Tag Description" on page 962.
Setting parameters:
<Data> block data
Binary block data with the following syntax:
#<Digits><Length><Binary data>
#
Indicates the start of the binary block
<Digits>
Decimal value
Gives the number of decimal digits used for the <Length>
value
<Length>
Decimal value
Number of bytes the follow in the <Binary data> part
<Binary data>
Binary data in ASCII format
Query parameters:
<Tag> 'comment' | 'copyright' | 'date' | 'lacpfilter' | 'marker name' |
'poweroffset'