Remote Control Commands
R&S
®
SMW200A
935User Manual 1175.6632.02 ─ 16
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SCONfiguration:PRESet
SCONfiguration:MODE ADV
SCONfiguration:FADing SISO4X1X1
SCONfiguration:BASeband:SOURce COUP
SCONfiguration:APPLy
ENTity1:SOURce1:BB:ARBitrary:COUPled:STATe 0
ENTity1:SOURce1:BB:ARBitrary:WAVeform:SELect "/var/user/lte"
ENTity1:SOURce2:BB:ARBitrary:WAVeform:SELect "/var/user/lte"
ENTity1:SOURce3:BB:ARBitrary:WAVeform:SELect "/var/user/3gpp"
ENTity1:SOURce4:BB:ARBitrary:WAVeform:SELect "/var/user/gsm"
ENTity1:SOURce4:BB:ARBitrary:COUPled:TRIGger:DELay:OFFSet 100
SOURce1:BB:ARBitrary:STATe 1
Manual operation:
See "Use one Waveform for all basebands" on page 308
[:SOURce<hw>]:BB:ARBitrary:COUPled:TRIGger:DELay:OFFSet <Offset>
Sets a time delay to delay the waveform processing of a particular baseband.
Parameters:
<Offset> float
Range: 0 to 2147483647/clockrate
Increment: 250E-12
*RST: 0
Example: see [:SOURce<hw>]:BB:ARBitrary:COUPled:STATe
on page 934
Manual operation: See "Delay" on page 309
[:SOURce<hw>]:BB:ARBitrary:WAVeform:CLOCk <Filename>, <Clock>
Appends information on the ARB clock rate to specified waveform file. The file must
contain I/Q and/or marker data and have been created with the command :MMEMory:
DATA:UNPRotected on page 783.
Parameters:
<Filename> string
Complete file path and file name with file extension (*.wv).
If the file is in the default directory, the file path can be omitted.
Setting parameters:
<Clock> float
Range: 400 to 100E6
Increment: 1E-3
*RST: 1E6