Remote Control Commands
R&S
®
SMW200A
941User Manual 1175.6632.02 ─ 16
Parameters:
<Clock> float
Increment: 1E-3
*RST: max SampleRate
Example: See Example "Creating a multi-segment waveform" on page 922
Manual operation: See "User Clock Rate" on page 336
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk:MODE <Mode>
Selects the clock rate mode for the multi segment waveform. Use the command [:
SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk to define the clock
in clock mode user.
Parameters:
<Mode> UNCHanged | HIGHest | USER
*RST: UNCHanged
Example: See Example "Creating a multi-segment waveform" on page 922
Manual operation: See "Clock" on page 336
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:COMMent <Comment>
Enters a comment for the selected configuration file.
Parameters:
<Comment> string
Example: See Example "Creating a multi-segment waveform" on page 922
Manual operation: See "Comment" on page 332
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:DELete <Filename>
Deletes the selected configuration file.
Setting parameters:
<Filename> string
Example: see Example "Creating a multi-segment waveform" on page 922
Usage: Setting only
Manual operation: See "Multi Segment Table, Append/Delete/Shift Seg."
on page 334
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:LEVel[:MODE] <Mode>
Selects the level mode, unchanged or equal RMS, for the multi-segment waveform.
SOURce Subsystem