Performing Measurements
R&S
®
ZNB/ZNBT
72Getting Started 1316.0062.02 ─ 26
You can also use the basic transmission test setup, e.g. if you want to measure
reflection and transmission parameters in parallel.
The analyzer provides special calibration types for reflection measurements. Use
the calibration wizard and select an appropriate type.
A full n-port calibration (TOSM, UOSM, TNA ...) corrects the system errors for all
transmission and reflection S-parameters.
Some of the trace formats are particularly suited for reflection measurements. For
instance, you can display the measured reflection coefficient S
11
in a Smith chart to
obtain the complex input impedance at port 1.
Reflection S-Parameter Measurement