Command Reference
R&S
®
ZNB/ZNBT
1028User Manual 1173.9163.02 ─ 54
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
Manual operation: See "Add Tr+Diag" on page 305
DISPlay[:WINDow<Wnd>]:CATalog?
Returns the numbers and names of all diagrams in the current recall set.
The response is a string containing a comma-separated list of diagram area numbers
and names, see example below. If all diagram areas have been deleted, the response
is an empty string ("").
Suffix:
<Wnd>
.
Number of a diagram. This suffix is ignored; the command
returns a list of all diagrams.
Example:
*RST; :DISP:WIND2:STAT ON
Create diagram no. 2.
DISP:WIND2:NAME 'S11 Test Diagram'
Assign a name to the new diagram.
DISP:CAT?
Query all diagrams and their names. As a default diagram no. 1
is created upon *RST, the response is ''1,1,2,S11 Test
Diagram'. The first diagram is not named; its default name is
equal to the diagram number.
CALC:PAR:SDEF 'Win2_Tr1', 'S11'
Create a trace named Win2_Tr1 to measure the input reflection
coefficient S
11
.
DISP:WIND2:TRAC9:FEED 'Win2_Tr1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC:CAT?
Query all traces in diagram area no. 2. The response is
'9,Win2_Tr1'.
Usage:
Query only
Manual operation: See "Title" on page 686
DISPlay[:WINDow<Wnd>]:MAXimize <Boolean>
Maximizes all diagram areas in the active recall set or restores the previous display
configuration.