EasyManua.ls Logo

Rohde & Schwarz ZNB - Page 1077

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1077User Manual 1173.9163.02 ─ 54
Manual operation: See "Add / Add All... / Replace / Apply / Apply to All"
on page 500
MMEMory:LOAD:CORRection:RESolve <Channel>[, <CalGroupFile>]
Resolves the pool link between channel <Channel> and a correction data set (cal
group file). After resolving the pool link, the analyzer keeps the previous system error
correction as a channel calibration ("Channel Cal"). A new calibration will replace the
channel calibration but not overwrite the old cal group file (and not affect other chan-
nels).
Setting parameters:
<Channel> Channel number of an existing channel. ALL resolves the pool
link for all channels.
<CalGroupFile> Optional string parameter to specify the name of the cal group
file. Cal group files must have the extension *.cal. The direc-
tory must not be specified; the analyzer always uses the default
path C:\Users\Public\Documents\Rohde-Schwarz\Vna
\Calibration\Data. If there is no link between <Ch> and the
specified file, the command has no effect. If no file is specified,
the command resolves any link between <Ch> and an arbitrary
cal group file.
Example: See MMEMory:LOAD:CORRection
Usage: Setting only
Manual operation: See "Resolve Pool Link / Remove Pool Link" on page 501
MMEMory:LOAD:CORRection:TCOefficient<Ch> <TraceFile>[, <Trace>]
Loads power meter (two-port) transmission coefficients from the specified power meter
correction file or trace file to channel <Ch>.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceFile> String parameter specifying the name and directory of the loa-
ded trace file. The R&S ZNB/ZNBT supports power meter cor-
rection list files (*.pmcl, generated using MMEMory:STORe:
CORRection:TCOefficient<Ch>), *csv, and Touchstone
(*.s1p, *.s2p, ...) files.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory
The file extensions *.s<n>p, *csv, and *.pmcl for Touch-
stone, ASCII, and power meter correction list files are manda-
tory.
*RST: n/a
SCPI Command Reference

Table of Contents

Other manuals for Rohde & Schwarz ZNB

Related product manuals