EasyManua.ls Logo

Rohde & Schwarz ZNB - Page 1105

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1105User Manual 1173.9163.02 ─ 54
The command fails unless the conditions for Touchstone file export are met; see "Con-
ditions for Touchstone file export" on page 144. Traces are created using the
CALCulate<Ch>:PARameter:SDEFine command.
Setting parameters:
<Channel> Channel number in the active recall set.
<TouchstoneFile> String parameter to specify the name and directory of the cre-
ated Touchstone file. The file extension *.s<n>p for a n-port
Touchstone file is mandatory.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory
<Format> COMPlex | LINPhase | LOGPhase
COMPlex - complex values (real and imaginary part).
LINPhase - Linear magnitude and phase.
LOGPhase - dB-magnitude and phase.
<ModeImpedance> CIMPedance | PIMPedance
CIMPedance: normalize to the common target impedance (from
options line); this is the default if the parameter is ommitted
PIMPedance: normalize to the individual port reference impe-
dances
see "Renormalization of S-parameters" on page 145
<Port>, <Port>, ... Physical port numbers
Example: Suppose that a full two-port calibration for ports 1 and 2 and
channel 1 has been performed, and that a DUT with two bal-
anced ports is connected. The analyzer measures an arbitrary
mixed mode S-parameter.
MMEM:STOR:TRAC:PORT 1, 'Test_CIMP.s2p',
COMPlex, CIMPedance, 1, 2
Calculate all single-ended S-parameters, renormalize them to
the common target impedance and store them to a two-port
Touchstone file.
MMEM:STOR:TRAC:PORT 1, 'Test_PIMP.s2p',
COMPlex, PIMPedance, 1, 2
Calculate all single-ended S-parameters, renormalize them to to
the individual port reference impedances and store them to a
two-port Touchstone file.
Usage:
Setting only
Manual operation: See "Save" on page 340
8.3.12 OUTPut Commands
The OUTPut... commands control the characteristics of the analyzer’s output ports.
SCPI Command Reference

Table of Contents

Other manuals for Rohde & Schwarz ZNB

Related product manuals