<Port1>, <Port2> Optional port restriction.
For a one-port standard the validity of the standard characteriza-
tion can be restricted to a single port, for a two-port standard it
can be restricted to a port pair (specified using ascending port
numbers).
Note: with a port restriction the defined standard becomes sex-
less. Hence, for each port (pair) there can be only one standard
of a given type, i.e. :SENSe:CORRection:CKIT:MOP
<ParameterList>,1 and :SENSe:CORRection:CKIT:FOP
<ParameterList>,1 will overwrite each other.
Parameters for setting and query:
<ConnType>,
<CalkitName>
String parameters uniquely identifying the cal kit to which the
standard belongs.
Note: If the specified cal kit does not exist, it is created with the
specified calibration standard.
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','New Kit 1'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
Manual operation: See "One Port Standards / Two Port Standards" on page 483
Table 8-12: Set command parameters
One port standards Two port standards
Open (Offset)
Short
Match Reflect Sliding
match
Symmetric
Network
Through,
Line
Attenuation
<Con-
nType>
mandatory
<CalKit-
Name>
mandatory (and must not be an empty string)
<Standar-
dLabel>
mandatory (but may be an empty string)
<MinFreq>,
<MaxFreq>
mandatory
<Delay-
Param>,
<Loss>,
<Z0>
mandatory mandatory mandatory mandatory not used mandatory not used
<C0>,
<C1>,
<C2>, <C3>
mandatory mandatory* mandatory* mandatory** mandatory** not used