Command Reference
R&S
®
ZNB/ZNBT
1157User Manual 1173.9163.02 ─ 54
Example:
CORR:COLL:CONN:PORTS SING
Select independent connector types at the ports.
CORR:COLL:CONN1 N50MALE; CONN4 N75FEMALE;
CONN2?
Select independent connector types at ports 1 and 4. The con-
nector type at port 2 is not changed; the query returns
N50FEMALE.
Manual operation: See "Same Connector All Ports / Same Gender All Ports"
on page 435
[SENSe<Ch>:]CORRection:COLLect:DELete
[<CalName>]
Deletes system error correction data generated and stored previously.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName> Name of the calibration (string parameter) defined together with
the calibration type ([SENSe<Ch>:]CORRection:COLLect:
METHod:DEFine).
ALL - the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy].
*RST: -
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the
active channel, then delete the data.
Usage: Setting only
Manual operation: See "Apply" on page 439
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel