Command Reference
R&S
®
ZNB/ZNBT
1177User Manual 1173.9163.02 ─ 54
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> OFF (0)
f
2
not used (default)
ON (1)
f
2
used
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 572
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe]
<Boolean>
Resets the offset parameters for all test ports to zero and the reference frequency to 1
GHz or queries whether any of the offset parameters are different from zero.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean> The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
ON - no effect
OFF - resets all length offsets to zero and the reference fre-
quency to 1 GHz
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST: OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation: See "Reset Offsets" on page 573
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe] <Boolean>
Toggles length/loss/fixture compensation for physical port <PhyPt> ON/OFF.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number