Command Reference
R&S
®
ZNB/ZNBT
1184User Manual 1173.9163.02 ─ 54
In MultiCal scenarios, use [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment:ALL:COUNt? to get the number of port assignments of all calibrations.
Suffix:
<Ch>
.
Channel number
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1186
Usage: Query only
Manual operation: See "Port Assignment (manual) " on page 422
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
<TestPort1>, <TestPort2>...
Creates the default port assignment(s) for the specified test ports.
The number of created assignments can be queried using the [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:COUNt? command.
Note that during the corresponding calibration sweep the R&S ZNB/ZNBT expects the
physical port connections to be established exactly as specified by the created port
assignments.
Use
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault on page 1185 to use auto-detection instead of defining "test port --> cal
unit port"-connections explicitely
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine or
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt to define (additional) port assignments manually
See Chapter 5.5.5.4, "Multiple Port Assignments", on page 173 for background infor-
mation.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1> Test port number 1
<TestPort2> Test port number 2
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1186
Usage: Setting only
Manual operation: See "Default Port Assignment" on page 423
SCPI Command Reference