Command Reference
R&S
®
ZNB/ZNBT
1189User Manual 1173.9163.02 ─ 54
[SENSe<Chn>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel <Chn>.
The response is a string variable containing the calibration state label in the trace list
('CAL', 'CAI', 'CAL OFF' ..; see Chapter 5.5.4, "Calibration Labels", on page 168).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See [SENSe<Ch>:]CORRection:DATE?
Usage: Query only
Manual operation: See "Apply" on page 439
8.3.14.9 [SENSe:]COUPle...
The [SENSe:]COUPle... commands select the sweep mode.
[SENSe<Ch>:]COUPle <Order>
Determines the order of partial measurements and sweeps.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the sweep mode applies
to all channels in the active recall set.
Parameters:
<Order> ALL | AUTO | NONE
AUTO - optimized display update: fast sweeps are performed in
alternated mode, slower sweeps in chopped mode
ALL - chopped sweep mode, complete all partial measurements
before proceeding to the next sweep point
NONE - alternated sweep mode on, reverse the order of partial
measurements and sweeps
*RST: NONE
Example:
COUP NONE
Activate the alternated sweep mode.
TRIG:LINK 'PPO'
Set the triggered measurement sequence equal to one partial
measurement. Each trigger event starts one partial measure-
ment for all sweep points.
Manual operation: See "Driving Mode" on page 542
8.3.14.10 [SENSe:]DC...
The [SENSe:]DC... commands select and configure the DC voltage measurement.
[SENSe<Ch>:]DC<DCInp>:RANGe <Voltage>
Assigns a DC voltage range to the rear panel input connector DC INPUT <DCInp>.
SCPI Command Reference