EasyManua.ls Logo

Rohde & Schwarz ZNB - Page 1195

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1195User Manual 1173.9163.02 ─ 54
Example: See SOURce<Ch>:FREQuency<PhyPt>:CONVersion:
ARBitrary:IFRequency
Manual operation: See "Load Match Correction" on page 497
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:FIXed<Stg> <InputOutput>
Selects the mixer input or output signal which is at the fixed frequency defined via
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:MFFixed.
Suffix:
<Ch>
.
Channel number.
<Stg> Counter for fixed signal (1 if only one mixer stage is used, 1 or 2
for 2 mixer stages).
Parameters:
<InputOutput> RF | LO | LO1 | LO2 | IF
RF - mixer input signal
LO | LO1 | LO2 - local oscillator signal 1 or 2 (for 2-stage mixer
measurements, see [SENSe<Ch>:]FREQuency:
CONVersion:MIXer:STAGes)
IF - mixer output signal (mixing product)
*RST: LO
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:FUNDamental <InputOutput>
Selects the mixer input or output signal which is at the fundamental frequency (channel
frequency, to be defined by means of [SENSe<Ch>:]FREQuency:STARt,
[SENSe<Ch>:]FREQuency:STOP, [SENSe<Ch>:]FREQuency:CW etc.).
Suffix:
<Ch>
.
Channel number.
Parameters:
<InputOutput> RF | LO | LO1 | LO2 | IF
RF - mixer input signal
LO | LO1 | LO2 - local oscillator signal 1 or 2 (for 2-stage mixer
measurements; see [SENSe<Ch>:]FREQuency:
CONVersion:MIXer:STAGes)
IF - mixer output signal (mixing product)
*RST: RF
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Frequency tab" on page 564