EasyManua.ls Logo

Rohde & Schwarz ZNB - Page 1196

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1196User Manual 1173.9163.02 ─ 54
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:IFPort <PortNumber>
Selects an analyzer port as receive port for the IF signal.
Suffix:
<Ch>
.
Channel number
Parameters:
<PortNumber> Range: 1 to number of test ports
*RST: 2
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Port selection" on page 564
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOMultiplier<Stg> <Numerator>,
<Denominator>
Selects the frequency conversion factors for the LO signal.
The conversion factor is a ratio of two positive integers: numerator and denominator.
Suffix:
<Ch>
.
Channel number
<Stg> Mixer stage signals LO 1 or LO 2; see [SENSe<Ch>:
]FREQuency:CONVersion:MIXer:STAGes
Parameters:
<Numerator>
Range: 1, 2, 3 ...
*RST: 1
<Denominator> Range: 1, 2, 3 ...
*RST: 1
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Multiplication tab" on page 565
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOPort<Stg> <SourceType>[,
<SourceNumber>]
Selects an analyzer or an external generator port as a signal source for the LO1 or
LO2 signal.
Note:
If External Switch Matrices are part of the RF connection configuration, operation with
more than one internal source is not supported. In this case only external generators
can be used as local oscillators.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference

Table of Contents

Other manuals for Rohde & Schwarz ZNB

Related product manuals