Command Reference
R&S
®
ZNB/ZNBT
1214User Manual 1173.9163.02 ─ 54
Example:
SENSe1:PAE:TYPE VOLTage
Select a measurement type with constant voltage source.
SENSe1:PAE:PARameters:U 5V; R 50Ohm
Define the parameters for the PAE measurement.
SENSe1:PAE:DCINput:MAIN DC1; :DC1:RANGe 20
Define the DC connector and power range
Alternative measurement type: constant current source.
SENSe1:PAE:TYPE CURRent
SENSe1:PAE:PARameters:I 100mA
SENSe1:PAE:DCINput:MAIN DC1; :DC1:RANGe 20
Alternative measurement type: voltange and current.
SENSe1:PAE:TYPE VCURrent
SENSe1:PAE:PARameters:R 50Ohm
SENSe1:PAE:DCINput:MAIN DC1; :DC1:RANGe 3
SENSe1:PAE:DCINput:SECondary DC2; :DC2:RANGe 20
For all measurement types:
CALCulate1:PARameter:SDEFine 'Trc1', 'PAE21'
DISPlay:WINDow1:TRACe:FEED 'Trc1'
Select a PAE measurement with port 1 as a source port, port 2
as a receive port.
Manual operation:
See "Measurement Type" on page 289
8.3.14.15 [SENSe:]PORT...
The [SENSe:]PORT... commands define the reference impedances at the physical
ports.
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence <RealPart>[, <ImaginaryPart>]
Specifies the complex reference impedance for the physical port numbered <PhyPt>
(impedance renormalization).
Use [SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe] to toggle between config-
ured and default reference impedances.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<RealPart> Real part of the port impedance.
Range: 1 mΩ to 10 MΩ
*RST: Default reference impedance of the connector fam-
ily assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
SCPI Command Reference