Command Reference
R&S
®
ZNB/ZNBT
1241User Manual 1173.9163.02 ─ 54
<Seg> Sweep segment number. This suffix is ignored; the command
returns the sum of all segments.
Example: See [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
Usage: Query only
Manual operation: See "Optional Columns" on page 405
8.3.14.19 [SENSe:]SWEep...
The [SENSe:]SWEep... commands provide general settings to control the sweep.
Most of the settings have an impact on the sweep time.
[SENSe:]SWEep:COUNt:ALL........................................................................................1241
[SENSe<Ch>:]SWEep:AXIS:FREQuency....................................................................... 1241
[SENSe<Ch>:]SWEep:AXIS:POWer.............................................................................. 1242
[SENSe<Ch>:]SWEep:COUNt.......................................................................................1243
[SENSe<Ch>:]SWEep:DETector:TIME........................................................................... 1243
[SENSe<Ch>:]SWEep:DWELl....................................................................................... 1244
[SENSe<Ch>:]SWEep:DWELl:IPOint............................................................................. 1245
[SENSe<Ch>:]SWEep:GENeration................................................................................ 1245
[SENSe<Ch>:]SWEep:GENeration:ANALog:CONDition?................................................. 1245
[SENSe<Ch>:]SWEep:POINts.......................................................................................1246
[SENSe<Ch>:]SWEep:STEP.........................................................................................1247
[SENSe<Ch>:]SWEep:TIME......................................................................................... 1247
[SENSe<Ch>:]SWEep:TIME:AUTO................................................................................1248
[SENSe<Ch>:]SWEep:TYPE.........................................................................................1249
[SENSe<Chn>:]SWEep:SRCPort...................................................................................1250
[SENSe:]SWEep:COUNt:ALL <Sweep>
Defines the number of sweeps to be measured in single sweep mode
(INITiate<Ch>:CONTinuous OFF). The setting is applied to all channels. Use
[SENSe<Ch>:]SWEep:COUNt to define the sweep count for a single channel.
Setting parameters:
<Sweep> Number of consecutive sweeps to be measured.
Range: 1 to 100000
*RST: 1
Example: See Chapter 9.2.4, "Data Handling", on page 1402.
Usage: Setting only
Manual operation: See "Sweeps" on page 412
[SENSe<Ch>:]SWEep:AXIS:FREQuency <Scale>
Selects either the channel base frequency or one of the port frequencies as the stimu-
lus axis in all diagrams of the active recall set. This command has no effect if no port-
specific frequencies are defined.