Command Reference
R&S
®
ZNB/ZNBT
1275User Manual 1173.9163.02 ─ 54
'GEN <n>' external generator port n (separate port no. by
blank, e.g. 'GEN 1')
*RST: 'COMBINER'
Example:
*RST; :SOURce:POWer:CORRection:IMODulation:PORT
'PORT 4'
Select the analyzer port no. 4 as a source port for the receiver
power calibration.
Manual operation: See "Src. for Rcvr. Cal" on page 554
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire] <SourceType>[,
<SourcePort>, <CalOnlyPortFreq>]
Selects the source for the source power calibration, starts and applies the source
power calibration.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number. This suffix is ignored because the port
number is specified in the parameter list.
Setting parameters:
<SourceType> PORT | GENerator | A1 | A2 | A3 | A4 | ESRC1 | ESRC2
PORT analyzer port is the source of the calibrated wave.
GENerator external generator is the source.
The port and generator numbers are specified by means of the
<SourcePort> parameter.
A1 | A2 | A3 | A4 direct analyzer port selection. The parame-
ters A1 and A2 are ZVR-compatible. No additional port number
parameter needs to be specified.
ESRC1 | ESRC2 direct external generator selection. The
parameters are ZVR-compatible. No additional parameters need
to be specified.
<SourcePort>
Analyzer port or generator port number, if PORT or GENerator is
the first parameter.
<CalOnlyPortFreq>
SCPI Command Reference