Command Reference
R&S
®
ZNB/ZNBT
1277User Manual 1173.9163.02 ─ 54
Example:
*RST; :SOURce:POWer:CORRection:NREadings 2
SOURce:POWer:CORRection:COLLect:AVERage:
NTOLerance 0.5
Increase the number of readings and reduce the power toler-
ance to improve the accuracy.
SOURce:POWer:CORRection:ACQuire PORT,1
Perform a source power calibration using port 1 as a source
port.
SOURce:POWer:CORRection:ACQuire:VERification:
RESult?
Query the result of the verification sweep. Possible response:
Ch1 1,9.9E+002
Usage:
Query only
Manual operation: See "Calibration Sweep Diagram" on page 445
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:AVERage[:COUNt]
<NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command: SOURce<Ch>:POWer<PhyPt>:CORRection:NREadings.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the limit
applies to all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the limit
applies to all sources.
Parameters:
<NoReadings> Number of readings
Range: 1 to 100
*RST: 2
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
(for equivalent command SOURce<Ch>:POWer<PhyPt>:
CORRection:NREadings)
Manual operation: See "Flatness Cal – Max Iterations" on page 491
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:AVERage:NTOLerance
<Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the toler-
ance value is valid for all sources.
SCPI Command Reference