Command Reference
R&S
®
ZNB/ZNBT
1309User Manual 1173.9163.02 ─ 54
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 420
SYSTem:COMMunicate:RDEVice:AKAL:DATE? <CalKitName>
Queries the creation date and time of the cal unit characterization (calkit file) <CalKit-
Name>. A possible response is 'Friday, May 26, 2011, 10:13:40'. An empty
string is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
Query parameters:
CalKitName String parameter containing the name of a cal unit characteriza-
tion (calkit file).
'Factory' denotes the factory characterization; an empty
string ' ' refers to the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 420
SYSTem:COMMunicate:RDEVice:AKAL:FRANge? <CalKitName>
Queries the frequency range of the cal unit characterization (calkit file) <CalKitName>.
The response consists of the start and stop frequencies in Hz, separated by a comma.
0,0 is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
Query parameters:
<CalKitName> String parameter containing the name of a cal unit characteriza-
tion (calkit file).
'Factory' denotes the factory characterization; an empty
string ' ' denotes the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 420