Command Reference
R&S
®
ZNB/ZNBT
1311User Manual 1173.9163.02 ─ 54
<Type> THRough | OPEN | SHORt | MATCh | MMTHrough |
MFTHrough | FFTHrough | MOPen | FOPen | MSHort | FSHort |
MMTCh | FMTCh
Standard types; for a description refer to table Standard types
and their parameters . The standard types of a particular charac-
terization can be queried via SYSTem:COMMunicate:
RDEVice:AKAL:CKIT:STANdard:CATalog?.
The factory characterization usually does not contain data for a
through standard; therefore a query of the type
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
'Factory', THR, S11, 1, 2 results in an error message.
<SParameter>
S11 | S12 | S21 | S22
S-parameter of the standard, use S11 for one-port standards,
one of the four for 2-port standards.
<FirstPort> First port number (sufficient for one-port standards). Port num-
bers can be omitted if the cal kit data is valid for all ports; see
MMEMory:LOAD:CKIT:SDATa.
<SecondPort> Second port number, for two-port standards. Port numbers can
be omitted if the cal kit data is valid for all ports.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 420
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Queries the warmup state of the connected calibration unit R&S ZV_Z5x. If several cal
units are USB-connected to the analyzer, the command queries the cal unit selected
via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
Possible responses are 1 (true, if the calibration unit has been connected for a suffi-
cient time to rach its operating temperature) or 0 (false). 0 is also returned if no calibra-
tion unit is connected.
Example:
See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 420
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:CATalog?
Queries the numbers of all configured external generators. The response is a string
containing a comma-separated list of generator numbers.