Command Reference
R&S
®
ZNB/ZNBT
1320User Manual 1173.9163.02 ─ 54
Example:
:INSTrument:SMATrix OFF
Remove all switch matrices from the RF configuration.
:SYSTem:COMMunicate:RDEVice:SMATrix:DELete
Unconfigure all switch matrices, i.e. remove them from the list of
configured devices.
:SYSTem:COMMunicate:RDEVice:SMATrix1:DEFine '',
'ZV-Z82-30', 'LAN', '192.168.0.42'
Register a single switch matrix of type ZV-Z82-30 (with two 2x5
submatrices).
:INSTrument:SMATrix ON
Add the matrix to the RF configuration, performing a default
assignment of VNA ports and test ports.
:SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
STARt
Start a manual RF connection configuration.
:SYSTem:COMMunicate:RDEVice:SMATrix1:CONFigure:
MVNA 1,1,3,2,2,3,4,4
Define non-default matrix-VNA port connections: matrix port 1 to
VNA port 2, matrix port 3 to VNA port 2, matrix port 2 to VNA
port 3 and matrix port 4 to VNA port 4 (default is
1,3,2,4,3,1,4,2).
:SYSTem:COMMunicate:RDEVice:SMATrix1:CONFigure:
MTESt 1,1,2,2,9,3,10,4
Configure 4 matrix test ports only: matrix test port 1 is test port
1, matrix test port 2 is test port 2, matrix test port 9 is test port 3
and matrix test port 10 is test port 4.
:SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
END
Apply the manual RF connection configuration.
Usage:
Event
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:TVNA <TestPort1>,
<VNAPort1>...
Sets/gets the test port connections of the VNA in a switch matrix RF connection setup,
i.e. the direct assignments of test ports to physical VNA ports.
Note: a manual RF connection configuration must be performed within a transaction,
enclosed by the commands SYSTem:COMMunicate:RDEVice:SMATrix:
CONFigure:STARt and SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
END. The modified configuration will not be applied until the end of this transaction.
Setting parameters:
<TestPort1> Number of the test port.
Test ports must be numbered subsequently starting at 1.
<VNAPort1> Number of the VNA port
Example:
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
TVNA 6,2,7,4
Test Ports 6 and 7 are assigned to VNA ports 2 and 4.