Command Reference
R&S
®
ZNB/ZNBT
1354User Manual 1173.9163.02 ─ 54
Parameters:
<Type> POINt | SWEep
ACQuisition
A trigger out signal is generated for every data acquisition phase
*RST: SWEep
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary<n>:OPOLarity <Type>
Selects the polarity of the output trigger pulses; see Figure 8-1.
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<Type> POSitive | NEGative
*RST: NEGative
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition <Type>
Specifies whether the trigger output pulse is sent before or after the selected sweep
interval (see TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval).
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<Type> BEFore | AFTer
BEFore
is appropriate if the device needs to be triggered before the data
is acquired (typical example: a power meter which is used for
power calibration)
AFTer
is appropriate if the device needs to be triggered just after data
acquisition (typical example: an external generator which must
be reconfigured to get ready for the next measurement)
*RST: AFTer
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
on page 1352