Command Reference
R&S
®
ZNB/ZNBT
1367User Manual 1173.9163.02 ─ 54
Setting parameters:
<CalKitFile> String parameter to specify the name and directory of the cal kit
file to be loaded.
Note: The loaded file must be a R&S ZVR-specific cal kit file
with the extension *.ck. VNA cal kit files (*.calkit) can be
imported using the MMEMory:LOAD:CKIT command. Agilent cal
kit files can be imported manually and converted into *.calkit
files.
Example:
CORR:CKIT:INST 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\ZCAN.ck'
Load the previously created R&S ZVR cal kit file ZCAN.ck from
the default cal kit directory.
MMEM:STOR:CKIT 'ZCAN', 'C:
\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\ZCAN.calkit'
Store the imported cal kit data to a VNA cal kit file
ZCAN.calkit (assuming that the cal kit name stored in
ZCAN.ck reads "ZCAN").
Usage: Setting only
[SENSe<Ch>:]CORRection:COLLect[:ACQuire] <Standard>[, <Dispersion>,
<Delay>]
Starts a calibration measurement in order to acquire measurement data for the
selected standards. The standards are reflection or transmission standards and must
be connected to port 1 or 2 of the analyzer.
Tip: Use the generalized command [SENSe<Ch>:]CORRection:COLLect[:
ACQuire]:SELected to obtain measurement data at arbitrary analyzer ports.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
R&S ZVR/ZVABT Compatible Commands