GUI Reference
R&S
®
ZNB/ZNBT
283User Manual 1173.9163.02 ─ 54
The "Imbal. CMRR" measurements require at least two logical ports are available, at
least one of them balanced.
Background information
Refer to the following sections:
Chapter 5.3.6, "Unbalance-Balance Conversion", on page 129
Chapter 5.3.6.3, "Imbalance and Common Mode Rejection", on page 132
6.2.8.1 Controls on the Imbal. CMRR Tab
The following buttons in the "Imbal. CMRR" tab open additional dialogs:
"Imbalance Differential...": see Chapter 6.2.8.2, "Imbalance Differential Dialog",
on page 284.
This button is only enabled, if a balanced and two single-ended logical ports are
available.
"Balanced Ports...": see Chapter 6.2.1.3, "Balanced Ports Dialog", on page 262
Imbalance/CMRR
Selects an imbalance or CMRR parameter as a measured quantity for the active trace.
These parameters are expressed as "Imb
<out><in>
" or "CMRR
<out><in>
", where <out> and
<in> denote the logical output and input port numbers of the DUT.
Remote command:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "IMB21" |
"IMB12" | "CMRR11" | "CMRR21" | "CMRR12" ...
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "IMB21" |
"IMB12" | "CMRR11" | "CMRR21" | "CMRR12" ...
Meas Softtool