567User Manual 1173.9163.02 ─ 54
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion
6.12.6.4 SMARTerCal (Cal Unit) Wizard for Mixer Mode
Allows to perfom a variant of an automatic "SMARTerCal" that is specially tailored to
the "Mixer Mode".
Access: CHANNEL – [CHANNEL CONFIG] > "Mixer Mode" > "SMARTerCal"
"Start... (Cal Unit)"
Offers similar functionality as the general SMARTerCal Wizard (Cal Unit), with the fol-
lowing differences:
Always all channels are calibrated
The affected test ports are already preselected and labeled according to their role
in the mixer measurement
The adequate calibration type is already preselected
A source flatness calibration can be automatically included as a the final calibration
step
Background information
Refer to Chapter 5.5.7, "SMARTerCal", on page 181 for background information
The wizard proceeds as follows:
1. Specify the remaining (non-fixed) properties of the automatic SMARTerCal
(required hardware, ports, connections etc.)