GUI Reference
R&S
®
ZNB/ZNBT
592User Manual 1173.9163.02 ─ 54
Figure 6-53: Fixture Modeling Dialog: ISD - Balanced Ports
Figure 6-54: Fixture Modeling Dialog: EZD - Single Ended Ports
The fixture modeling proceeds in the following steps:
1. Measure one or more test coupons for the related fixture; see Chapter 6.13.5.2,
"Measure Coupon", on page 594
The tools differ in the available coupon types.
2. ISD and SFD only: measure the DUT with the fixture; see Chapter 6.13.5.3, "Mea-
sure DUT + Test Fixture", on page 595
3. Run the selected tool to calculate the Touchstone files modeling the test fixture;
see Chapter 6.13.5.4, "Save Fixture Model", on page 596.
The calculated sNp file(s) can then be used to deembed the DUT at selected
port(s).
Offset Embed Softtool