Command Reference
R&S
®
ZNB/ZNBT
782User Manual 1173.9163.02 ─ 54
MMEMory Commands........................................................................................ 1063
OUTPut Commands............................................................................................1105
PROGram Commands........................................................................................ 1110
[SENSe:] Commands.......................................................................................... 1113
SOURce Commands...........................................................................................1251
STATus Commands............................................................................................ 1299
SYSTem Commands...........................................................................................1302
TRACe Commands.............................................................................................1345
TRIGger Commands...........................................................................................1348
8.3.1 CALCulate Commands
The CALCulate... commands perform post-acquisition data processing. Functions
in the SENSe subsystem are related to data acquisition, while the CALCulate subsys-
tem operates on the data acquired by a SENSe function.
8.3.1.1 CALCulate:CLIMits...
The CALCulate:CLIMits... commands control the composite limit check.
CALCulate:CLIMits:FAIL?
Returns a 0 or 1 to indicate whether or not a global, composite limit check on several
traces has failed.
Since V2.20 of the R&S ZNB/ZNBT FW the result is automatically recalculated when-
ever a relevant setting is changed, i.e. a subsequent query will return the updated limit
violation state.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:CLIM:FAIL?
Query the result for the composite limit check. As only one trace
is tested, the response should be equal to the previous
response.
Usage: Query only
Manual operation: See "Global Check" on page 345
8.3.1.2 CALCulate:DATA...
The CALCulate:DATA... commands provide access to the results of a measure-
ment.
SCPI Command Reference