Command Reference
R&S
®
ZNB/ZNBT
821User Manual 1173.9163.02 ─ 54
Example:
*RST; :CALCulate1:TRANsform:TIME:STATe ON
CALCulate1:FILTer:GATE:TIME:STATe ON; SHOW ON
Activate time domain representation and a time gate in channel
no. 1. Display the time gate
CALCulate1:FILTer:GATE:TIME:STARt 2ns; STOP 3
ns
Restrict the time gate to the time interval between 2 ns and 3 ns.
CALCulate:FILTer:GATE:TIME:AOFFset ON
Activate an offset of the time gate according to a new delay set-
ting.
SENSe1:CORRection:EDELay1:TIME 1ns
Specify a 1 ns delay at port 1.
CALCulate1:FILTer:GATE:TIME:STARt?; STOP?
Query the time gate position. The response is 1E-009;2E-009.
Manual operation:
See "Adjust Time Gate" on page 582
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE] <TimeGateFilter>
Selects the time gate filter type, defining what occurs to the data in the specific time
region.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGateFilter> BPASs | NOTCh
BPASs
Band pass filter: Pass all information in specified time region and
reject everything else
NOTCh
Notch filter: Reject all information in specified time region and
pass everything else
*RST: BPASs
Example:
*RST; :CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time gate.
CALC:FILT:TIME NOTCh
Select a notch filter in order to reject unwanted pulses.
Manual operation: See "Bandpass / Notch" on page 324
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer <CenterTime>
Defines the center time of the time gate.
Suffix:
<Chn>
.
Channel number used to identify the active trace