Command Reference
R&S
®
ZNB/ZNBT
823User Manual 1173.9163.02 ─ 54
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGate> MAXimum | WIDE | NORMal | MINimum
MINimum - Steepest edges (rectangle)
WIDE - Normal gate (Hann)
NORM - Steep edges (Hamming)
Maximum - Maximum flatness (Bohman)
*RST: WIDE
Example:
*RST; :CALC:FILT:TIME:SHAP?
Reset the instrument and query the type of time gate used. The
response is WIDE.
Manual operation: See "Shape" on page 325
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW
<Boolean>
Enables or disables permanent display of the gate limits.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON - time gate permanently displayed
OFF - time gate hidden
*RST: OFF
Example: See CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
Manual operation: See "Show Range Lines" on page 324
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN <Span>
Defines the span of the time gate.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Span> Span of the time gate.
Range: 2E-012 s to 200 s
Increment: 0.1 ns
*RST: 5E-009 s
Default unit: s
Example: See CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
Manual operation: See "Gate Span" on page 321