Command Reference
R&S
®
ZNB/ZNBT
837User Manual 1173.9163.02 ─ 54
CALCulate:FMODel:SFD<Ph_pt>:RUN:RUN
Runs the SFD tool.
Before executing this command, make sure that:
The test coupon measurement (using CALCulate:FMODel:SFD<Ph_pt>:
COUPon:MEASure) finished successfully, or the test coupon data were success-
fully loaded from file (using CALCulate:FMODel:SFD<Ph_pt>:COUPon:
MEASure:FILename
The measurement of DUT + test fixture (using CALCulate:FMODel:
SFD<Ph_pt>:DUT:MEASure) finished successfully.
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\Vna\Embedding. If result files with the same name already exist, they are
overwritten.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Event
Manual operation: See "Run <Fixture Modeling Tool>" on page 597
CALCulate:FMODel:SFD<Ph_pt>:RUN[:STATe] <Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the deembedding files shall be assigned after the tool has run.
Suffix:
<Ph_pt>
.
Physical port number for single-ended deembedding, logical port
number for balanced deembedding.
Parameters:
<Boolean> 1 (ON) if a deembedding file shall be assinged to port <Pt>, 0
(OFF) otherwise
Manual operation: See "Apply" on page 597
CALCulate:FMODel:SFD<Ph_pt>:TOTaldiffcfg <SFDPortConfig>
Tells the SFD tool about the port ordering of theTest Fixture.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<SFDPortConfig> ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
Manual operation: See "Total Port Ordering" on page 602