CALCulate:FMODel:EZD<Ph_pt>:RUN:RUN
Runs the EZD tool.
Before executing this command, make sure that:
The test coupon measurement (using CALCulate:FMODel:EZD<Ph_pt>:
COUPon:MEASure) finished successfully, or the test coupon data were success-
fully loaded from file (using CALCulate:FMODel:EZD<Ph_pt>:COUPon:
MEASure:FILename
The measurement of DUT + test fixture (using CALCulate:FMODel:
EZD<Ph_pt>:COUPon:MEASure) finished successfully.
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\Vna\Embedding. If result files with the same name already exist, they are
overwritten.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Event
Options: R&S ZNB/ZNBT-K210
Manual operation: See "Run <Fixture Modeling Tool>" on page 597
CALCulate:FMODel:EZD<Ph_pt>:RUN[:STATe] <Boolean>
For a fixture modeling with the EZD tool, this command allows to specify the ports to
which the deembedding files shall be assigned after the tool has run.
Suffix:
<Ph_pt>
.
Physical port number for single-ended deembedding, logical port
number for balanced deembedding.
Parameters:
<Boolean> 1 (ON) , 0 (OFF) otherwise
ON (1)
if a deembedding file shall be assinged to port <Ph_pt>
OFF (0)
otherwise
Options: R&S ZNB/ZNBT-K210
Manual operation: See "Apply" on page 597
More...
CALCulate:FMODel:REName <Boolean>
If set to ON (1), the names of subsequently generated "Test Coupon" and "DUT + Test
Fixture" files are prefixed with the current date and time.
Parameters:
<Boolean>