CALCulate<Chn>:RIPPle:FAIL?.......................................................................................912
CALCulate<Chn>:RIPPle:RDOMain:FORMat....................................................................912
CALCulate<Chn>:RIPPle:SEGMent:COUNt?....................................................................913
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]........................................................... 913
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit............................................................... 914
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?...........................................................914
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STARt...............................................915
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STOP............................................... 915
CALCulate<Chn>:RIPPle:SOUNd[:STATe]........................................................................916
CALCulate<Chn>:RIPPle:STATe......................................................................................916
CALCulate<Chn>:RIPPle:STATe:AREA............................................................................ 916
CALCulate:RIPPle:DISPlay:RESult:ALL[:STATe] <Enable>
Configures the display of ripple check info fields for the active recall set.
Parameters:
<Enable> ON - Info fields are displayed for all traces, for which a limit
check is enabled.
OFF - Only the info field for the active trace is displayed (if the
ripple check is enabled for this trace).
*RST: OFF
Manual operation: See "Show Results All Traces" on page 353
CALCulate:RIPPle:FAIL:ALL? [<RecallSet>]
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed for at
least one channel in the referenced recall set.
Query parameters:
<RecallSet> Recall set name; if omitted the active recall set is used
Usage: Query only
Manual operation: See "Ripple Check" on page 352
CALCulate<Chn>:RIPPle:CLEar
Resets the limit check results for the ripple test.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage: Event
Manual operation: See "Clear Test" on page 354