Trying Out the Instrument
R&S
®
ZNLE
41Getting Started 1323.2873.02 ─ 08
1. Connect the DUT between test ports 1 and 2 of the network analyzer as
shown above.
2. Switch on the instrument and start the VNA application.
Proceed as described in Chapter 3.6, "Switching On or Off", on page 16.
3. Use the [Preset] key to restore a well-defined instrument state.
The analyzer is now set to its default state. The default measured quantity is
the transmission S-parameter S
21
.
Select [Trace] and use the control elements in the "Traces" softtool tab if
you wish to create additional traces and diagrams.
5.1.1.2 Selecting the Sweep Range and Other Parameters
After a system preset the display shows a diagram with a dB magnitude scale,
and the S-parameter S
21
is selected as a measured quantity. This S-parameter is
the forward transmission coefficient of the DUT. It is defined as the ratio of the
Performing Measurements