Instrument Construction and Function Description R&S ZVB
1145.1078.82 3.66 E-4
Group 1:
General Functions
Functions Servicefunction Data Serv. Lev.
Enables/disables the peak detector and RMS
detector independently of the model
1.0.0.1.X
0 Use disabled
1 Use enabled
2
Enable/disables the arbitrary mode settings in
the port configuration (e.g. LowNoise/
LowDistortion) independently of the model
1.0.0.5.X
0 Use disabled
1 Use enabled
2
Activates or deactivates the setting of measured
values to default values when status messages
are issued due to HW faults
1.1.0.2.X
0 Deactivate default values
1 Activate default values
0
Activates or deactivates factory calibration for
the active setup
1.1.0.3.X
0 Deactivate factory calibration
1 Activate factory calibration
0
Group 2:
General Functions
Functions Service function Data Serv. Lev.
Suppress error-message box 2.0.0.0.X
X = 0 MSG box is output (default state)
X = 1 MSG box is suppressed (error is
nevertheless entered in the log file)
1
Selftest all boards 2.0.0.5.0.0 0, 1 or 2
Network controller
Functions Service function Data Serv. lev.
ND1:Read Temp (addr.:68) 2.1.2.3 1
ND2:Read Temp (addr.:68) 2.2.2.3 1
ND[1..2] Selftest 2.[1-2].2.5.0.0 0, 1 or 2
NA1:Read Temp (addr.:69) 2.1.3.3 1
NA2:Read Temp (addr.:69) 2.2.3.3 1