Manual Alignment and Recording Correction Values R&S ZVB
1145.1078.82 2.4 E-4
Manual Alignment and Recording Correction Values
In the sequel, the test equipment and the instrument preparations required to manually align the ZVB
and each of the alignments are described.
Preliminary remarks
The analyzer must be allowed to warm up for at least 30 minutes before alignment. This is the only way
of ensuring that the guaranteed data are met.
Test Equipment
Table 2-1 Test equipment for manually aligning the R&S ZVB
Item Type of equipment Recommended
specifications
Recommended
model
R&S Order No. Application
1 Spectrum analyzer Counter mode:
Min. resolution: 100 Hz
Max. rel. frequency
deviation: 10
-6
R&S FSU 26 1166.1660.26 Frequency accuracy of
the reference oscillator
2 Signal generator 300 kHz to 20 GHz R&S SML01
R&S SMR20
with option R&S
SMR-B15
1090.3000.11
1104.0002.20
1104.4989.02
Recording correction
values
3 Power meter 300 kHz to 20 GHz R&S NRVD 0857.8008.02
Recording correction
values
4 Power sensor 300 kHz to 8 GHz R&S NRV-Z51 0857.9004.02 Recording correction
values on R&S ZVB4/8
4 Power sensor 10 MHz to 20 GHz R&S NRV-Z52 0857.9204.02 Recording correction
values on R&S ZVB20
5 Power splitter N, 50 Ohm,
eq
< 0.05 (50 MHz to 8
GHz)
Output tracking < 0.15 dB
Weinschel 1870A
- Recording correction
values on R&S ZVB4/8
5 Power splitter 3.5mm, 50 Ohm,
Output tracking < 0.25 dB
Weinschel 1593 - Recording correction
values on R&S ZVB20
6 PC with
GPIB-Interface
Pentium, WinXP, WinNT
GR AT-GPIB IEEE4888 IF
PCI National Instruments
NI-488 PCI-GPIB
1072.6101.00
Recording correction
values
7 Alignment Software R&S ZVAB-Service 1302.4460.00 Recording correction
values
8 Power supply 2x 0 to 10 V Aligning the DC inputs
9 DC meter R&S URE 0350.5315.02 Aligning the DC inputs