Instrument Construction and Function Description R&S ZVB
1145.1078.82 3.24 E-4
Putting into operation
Connect the mains cable and turn on at the power on switch. The instrument is now in standby mode.
Insert the BIOS disk in the floppy disk drive.
Turn on the instrument and wait for the first beep. Press the DEL key. The instrument should now
display the setup menu.
Select Advanced BIOS Features
Enter
Select First Boot Device
Select Floppy using page up/down key
Press F10 key (save)
Enter
BIOS has now been programmed.
Do not turn the instrument off when the program is running.
Follow the instructions on the screen.
Select Service Level 2 (see Service Functions).
Check the protocol file for errors:
[ INFO : Error Log ]