R&S ZVL Measurement Examples
Basic Measurement Tasks
Operating Manual 1303.6580.32-05
107
To Set up a Segmented Frequency Sweep...
Press the Sweep key or click Channel - Sweep to access the Sweep submenu.
Click Sweep Type - Define Segments to call up the dialog defining the individual sweep segments.
Proceed as described in Define Segments to customize your sweep range. Make sure to define at
least 2 distinct sweep frequencies.
Use the controls in the Individual Settings panel to define the internal generator power, measurement
(IF) bandwidth and other measurement settings for each sweep segment.
Close the Define Segments dialog.
Click Sweep Type - Segmented Frequency.
If the active diagram is a Cartesian diagram, the x-axis is linearly scaled and ranges from the lowest to the
highest frequency in all segments.