EasyManua.ls Logo

Rohde & Schwarz ZVL series - Page 178

Rohde & Schwarz ZVL series
624 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ZVL GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
163
Target Search activates the search for the target value.
Target Search > activates the search for the target value to the right of the active marker.
< Target Search activates the search for the target value to the left of the active marker.
Def Value specifies the target value.
Search Range confines the search to a subrange of the sweep.
Target Search
Activates the search and sets the active marker to the defined target value. If the active trace contains no
markers, a marker Mkr 1 is created to indicate the search result.
If the target value occurs at several stimulus values, the marker is placed to the search result with the
smallest stimulus value. The other measurement points with the same target value can be located using
the Target Search > function.
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active
marker is not moved away from its original position.
Remote control:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute TARGet
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Target Search >
Activates the search to the right of the active marker position and sets the active marker to the defined
target value. The target search range is between the active marker position and the end (Stop) of the
sweep range. If the active trace contains no markers, a marker Mkr 1 is created to indicate the search
result, and the target search range starts at the beginning (Start) of the sweep range.
If the target value occurs at several stimulus values, the marker is placed to the search result with the
smallest stimulus value. The other measurement points with the same target value can be located using
Target Search > repeatedly.
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active
marker is not moved away from its original position.
Remote control:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute RTARget
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
< Target Search
Activates the search to the left of the active marker position and sets the active marker to the defined