R&S ZVL GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
184
F
ull One-Port P2 Cal o
pens the calibration wizard to perform a full one-port calibration at port no.
2 and re-adjust the reference plane (distance zero).
Fault Limit defines a minimum response value in dB for a trace maximum to be considered as
being due to a fault.
Show Fault List displays a list of all peaks that fulfil the Fault Limit condition.
Export Fault List opens a Save As... dialog to write the fault list data to an ASCII file.
Distance-to-Fault
Enables or disables the location of a discontinuity on a transmission line (with option R&S ZVL-K2). The
distance-to-fault measurement requires a linear frequency sweep (Channel – Sweep – Sweep Type – Lin.
Frequency). The analyzer automatically disables the distance-to-fault measurement as soon as a different
sweep type is selected.
Default settings for distance-to-fault
The following settings are performed automatically when a distance-to-fault measurement is enabled
(Trace – Meas– Distance-to-Fault: On):
The reflection coefficient S22 is selected as measured quantity.
The stimulus axis shows a distance window. The start and stop distance is displayed below the
channel settings.
The cable type (e.g. Ideal Air Line) is displayed in the trace list. A context menu opens the Cable
Type dialog.
The number of points is increased, if necessary. See Auto Number of Points.
Remote control:
CALCulate<Chn>:TRANsform:DTFault:STATe
Start Distance..., Stop Distance...
Defines the distance range (x-axis range) for the distance-to-fault measurement. The Start Distance and
the Stop Distance correspond to the left and right edge of the diagram area; they are shown below the
channel settings:
Distance trace settings
While a distance-to-fault measurement is active the trace settings behave as follows:
Start Distance... and Stop Distance... configure the distance axis.
dB Mag is the recommended Trace Format, although other formats are available.
Limit lines can be defined like the limit lines for time sweeps.
The marker and bandfilter search functions are available for the transformed trace.
Note that only the reflection coefficient S
22
(and the corresponding impedance and admittance
parameters) are available as measured quantities.
Remote control:
CALCulate<Chn>:TRANsform:DTFault:STARt
CA
LC
ulate<Chn>:TRANsform:DTFault:STOP
CALCulate<Chn>:TRANsform:DTFault:CENTer
CALCulate<Chn>:TRANsform:DTFault:SPAN