R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
219
Start Cal opens a submenu to select a new calibration and start the calibration wizard.
Repeat Prev Cal reopens the wizard to repeat and optimize the previous calibration.
Correction Off activates or deactivates the system error correction in the active channel.
Port Extensions provides a selection of length offset parameters to shift the measurement plane.
Cal Manager opens a dialog to store system error correction data to a Cal Pool and to assign
correction data to channels.
Recall Last Cal Set loads and activates the setup for which the last calibration was performed.
Cal Kits opens a dialog to manage the calibration kits in use, add new kits and import or export
kits.
Start Cal
The Start Cal submenu selects the calibrated ports and the calibration type.
The calibration types depend on the number of test ports of the analyzer. For a four-port unit:
One-Port P1 and One-Port P2 open submenus to select a one-port calibration at test ports PORT
1 and PORT 2, respectively.
Two-Port P1 P2 opens a submenu to select a two-port calibration at test ports PORT 1 and
PORT 2.
Calibrations can be performed automatically using the Calibration Unit (accessory R&S ZV-Z5x).
Remote control:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
One-Port P1, One-Port P2
Opens a submenu to select a one-port calibration at test ports PORT 1 or PORT 2. The two submenus are
identical: